Melexis’s 1.8 mm infrared sensor has to look past the watch housing

Illustrative image: Jens Mahnke on Pexels
Melexis can fit its new infrared temperature sensor in a package measuring 1.8 × 1.8 × 0.75 mm, but the amount of room it needs inside a watch depends on where it sits. Recess the MLX90636 behind a narrow opening and the housing can encroach on its 50° field of view. Move it further from the wrist and it sees a larger patch of skin. The package may be tiny, yet the opening above it and the distance to the skin still decide what reaches the thermopile. Any protective window has to pass the infrared wavelengths used for the measurement.
The part is factory calibrated to a stated ±0.2 °C across the human body temperature range. Even if a watch preserves that accuracy in the assembled device, the reading describes the surface it sees. The wrist can stay cold after the wearer steps indoors, while charging or radio activity can raise the temperature inside the watch. Both affect the measurement problem, but at different points: one changes the target and the other changes the conditions around the sensor.
Melexis is researching the next step, a model that estimates core temperature from wrist measurements. The interesting validation would compare the model against a suitable core-temperature reference across wearers and changes in fit and ambient temperature. In particular, a test should distinguish the sensor drifting from the model getting a plausible skin reading and drawing the wrong conclusion from it. The component specification has no way to settle that distinction.
Sources: Melexis announcement and wrist-to-core research; MLX90636 product specifications